Linear actuator for pipette travel: what decides the choice

Linear motion, bearings and guides Published 8 min read By AutomateDirectUSA engineering team

Short answer

A guided stepper-driven lead-screw axis with its own RS-232 or RS-422 motion controller is the practical choice for a 300 g pipette over about 150 mm of travel. The decision turns on what the 2 mm measures at the pipette tip, the offset moment on the carriage guide, power-off holding on a vertical mount, and a controller that stores programmed positions.

On this page
  1. Decide what the 2 mm measures at the tip
  2. The load is light; the moment and the holding decide
  3. Worked example: the static weight of a 0.300 kg pipette on a 150 mm vertical stroke
  4. Candidate architectures
  5. The serial port is not the motion language
  6. The decision for each case
  7. Questions asked

Use a guided electric axis: a stepper motor turning a lead screw inside a supported carriage, driven by a controller that stores your move sequence and takes commands over RS-232 or RS-422. For a 300 g pipette on about 150 mm of vertical travel, with positions held to roughly 2 mm, that architecture clears every number in your specification, and payload is not what decides the model. The decision turns on four things: what that accuracy figure measures at the pipette tip, the offset load moment on the carriage guide, what holds the carriage when power is off, and whether the controller accepts programmed positions rather than raw motor pulses. Work those four and each standard actuator style either passes or fails cleanly.

Decide what the 2 mm measures at the tip #

Three specifications get called accuracy, and they fail in different ways. Resolution is the smallest increment you can command. Repeatability is the spread of arrivals when the same point is commanded again and again. Accuracy is the gap between where the axis stops and where that point physically is. Command resolution and achieved position come apart exactly as in servo positioning resolution versus stepper step angle, and only repeatability and accuracy can be measured at your pipette.

Resolution is not your problem. A 1.8-degree stepper makes 360° ÷ 1.8° = 200 full steps per revolution; turning a 5 mm-lead screw, a typical pairing, that is 5 mm ÷ 200 = 25 µm of travel per full step, eighty times finer than your band. What matters is the tip, not the carriage: the carriage can land exactly on its coordinate while the pipette misses, because the holder bends, the pipette sits at an angle, or the frame shifts between the axis and the tubes. Reference every position at the tip, against the actual tubes.

Two definitions fix the requirement before any part is chosen. First, decide whether the band applies at a few working heights or at any height in the stroke: a few heights need repeatability, any height needs accuracy plus a reference procedure. Second, decide the approach direction: a height approached from the same side every time never shows the screw's backlash, while a height approached from both sides shows it twice over. One-direction approach is the cheapest accuracy you will ever buy.

For this duty the commitment is easy. A guided lead-screw axis, positions referenced at the pipette tip and every move approached from one direction, sits far inside a 2 mm band. The residual risk is the holder and the frame between carriage and tubes, not the actuator.

The load is light; the moment and the holding decide #

The axial load is a few newtons — the worked example below puts the number on it — so unlike sizing a long-stroke linear actuator for a 500 N push, thrust is not the specification that picks the part. Three other things are.

The offset moment. The pipette hangs outboard of the carriage, so its weight acts at a perpendicular offset from the guide: the moment equals weight times offset, in newton-metres. Read the permitted load moments for the exact axis in its datasheet — they depend on frame size and on whether the axis stands vertical or horizontal — because a tiny axial load does not excuse a long overhang. If the holder cannot come inboard, run the pipette on a second round shaft carried by linear bearings and bushings.

The upward force balance. On an upward move the drive supplies the total moving mass times the sum of gravity and acceleration, plus guide friction, plus any process force such as a plunger push: F_up = m × (g + a) + F_friction + F_process, with force in newtons, mass in kilograms and acceleration in metres per second squared. At pipette masses the acceleration term is exactly as large as your motion ramps make it, and modest ramps keep it small.

The power-off behaviour. A vertical axis is a gravity load, and a thrust rating says nothing about holding it: vertical applications need a load-holding design as well as a thrust figure. Neither screw friction nor an energized stepper's holding torque retains the carriage when power is cut — a de-energized stepper keeps only a small detent torque. Choose one of three answers before ordering: a power-off brake on the motor, a counterbalance such as a spring or gas strut taking part of the weight, or a mechanical rest stop the carriage settles onto.

One safety note: a vertical carriage with a tool on it is a falling-load hazard, so write the power-off behaviour into the machine's risk assessment — ISO 12100 requires hazards and their risk-reduction measures to be documented there — the same stored-energy discipline as stopping a vertical air cylinder on a light curtain trip.

Worked example: the static weight of a 0.300 kg pipette on a 150 mm vertical stroke #

Inputs, with where each comes from:

  • Pipette mass, m = 0.300 kg: your stated 300 g pipette.
  • Gravitational acceleration, g = 9.81 m/s²: the standard value.
  • Holder and carriage attachments: not in your numbers; weigh them and add them before closing the check below.

Rule, in words: the static weight of a mass is the mass multiplied by the acceleration of gravity.

Arithmetic:

F = m × g = 0.300 kg × 9.81 m/s² = 2.943 kg·m/s² ≈ 2.94 N

Result: with the axis vertical, the pipette alone presses down the screw with about 2.94 N. That is a calculated example for your stated inputs, not a measurement.

Decision, with its threshold: by rule of thumb, while the total moving weight — pipette, holder and carriage attachments together — stays under about 10 N, axial thrust does not decide a guided screw-axis selection; the permitted load moment and the power-off holding do. So weigh the holder: under that figure, the force question is closed and the effort goes into the moment and holding checks. Over it, redo the upward balance with your real acceleration and friction before choosing a frame size.

Candidate architectures #

The shortlist for a light, programmed, serial-commanded vertical move:

Architecture Why it is in the running What decides it
Guided stepper-driven lead-screw axis with its own controller Programmed intermediate heights are what the controller is built for, and the stepper holds position while energized Torque at move speed, permitted load moments, one-direction approach, power-off holding
Integrated unit with motor, screw, drive and controller in one housing Fewest parts to wire; stored positions over a serial line are its designed use Documented command set, usable travel, external guidance for an overhung holder, repair path
Rodless positioning actuator, belt or screw driven A carriage plate suits mounting a pipette holder directly Load-moment ratings, transmission type, controller compatibility
Pneumatic cylinder with a directional valve Robust and simple for endpoint-to-endpoint motion Fails the programmed-positions requirement unless the sequence is two fixed stops

Size the stepper on the torque it delivers at your move speed, not the holding torque on the label — torque falls as speed rises, and whether a stepper has enough of it at speed is the same question as stepper versus servo torque for robot joints. Two open-loop truths to design around: a stalled stepper loses steps silently and the controller cannot tell, and a de-energized stepper holds almost nothing, which is why the power-off decision above exists. Home the axis at every power-up so the controller's count and the physical position start from the same place.

The serial port is not the motion language #

RS-232 names an electrical interface and nothing more. It does not make a motor programmable: the host must talk to a controller that understands a sequence — move to height, wait, move to the next height — and turns it into motor phases. A bare actuator with a serial connector and no controller behind it fails your requirement before the first byte.

RS-422, which integrated motor-screw-controller units often use, is a different electrical standard: differential signalling over a twisted pair instead of the single-ended levels of RS-232. Either one carries motion commands. Match the interface to your host, or put a proper level converter between them; a rewired cable does not translate signals, so take the wiring from both devices' documentation.

From the exact controller's communication manual, confirm four things before buying: the electrical interface and connector pinout; the serial settings — baud rate, data bits, parity; whether it takes individual position commands, stores a sequence internally so the host sends one trigger per cycle, or needs the host to coordinate every move; and how the host reads back arrival and fault status. Your series of steps wants stored sequences or individual commands, and stored is the cleaner fit. Which host runs the show — the actuator's own controller, a PLC or an industrial PC — is an application-fit decision, the same reasoning as application fit deciding PLC selection rather than the programming language.

The decision for each case #

  • A few programmed heights, each held inside the band: a guided stepper-driven lead-screw axis with a controller that stores the sequence, positions referenced at the pipette tip, every move approached from one direction. This is the default answer to your specification.
  • Calibrated absolute heights anywhere in the stroke: the same axis plus its documented accuracy figure and a one-time tip calibration against a reference height, re-homed at every power-up before the first move.
  • A sequence that is truly two fixed endpoints: a pneumatic cylinder with hard stops meets the positioning need but not the serial programmability, so stay electric unless the sequence will never change.
  • Vertical mounting: fit the power-off answer — brake, counterbalance or rest stop — and write it into the risk assessment; mounting horizontally removes most of the problem when the process geometry allows.
  • The interface: RS-232 and RS-422 both do the job; choose on the controller's documented command set, convert levels with a real converter where the two standards meet, and bench-test a stored sequence end to end before the axis goes into the cell.

Questions asked

Do I need an encoder to hold the position band with a stepper axis?

No. An open-loop stepper on a lead screw already resolves tens of micrometres per step, so an encoder adds nothing to resolution. It earns its place only if a silently lost step is unacceptable: with an encoder the controller can compare commanded and achieved position and raise a fault instead of carrying on. The cheaper alternative is to re-home the axis at every power-up and after any stall or manual move.

How do I prove the tip position before wiring the cell?

Bench it in its final orientation. Command the lowest and highest working height ten times each, first always from the same direction, then from both, and measure the pipette tip against a gauge block or a fixed mark at every arrival. The spread of the readings is your measured repeatability; the offset of their centre from the target is your accuracy error. Ten bidirectional arrivals inside the band is evidence no datasheet line gives you.

Can I mount the axis horizontally instead?

Usually yes, and it removes most of the holding problem: gravity then loads the guide sideways instead of pulling the carriage down the screw, so power-off retention stops being a selection criterion. The pipette still has to dip vertically, which means an L-shaped holder or a bell-crank, and that holder grows the offset moment on the carriage. A small brake on a vertical axis is often simpler than the linkage a horizontal mount needs.

Keep reading

  • Selection guide

    Sizing a Long-Stroke Linear Actuator for a 500 N Push

    Measure force versus position through the seal before choosing hardware, then size from F_required = F_resistance + m_total × a. With 500 N and an 80 kg workpiece the workpiece-only acceleration ceiling is 6.25 m/s². Order an axis rated at least twice the measured requirement — 1000 N against that process force — and let the thrust-speed envelope, the screw-speed limit over the stroke, and the duty cycle pick belt, ball screw or worm drive.

    Linear motion, bearings and guides

  • Selection guide

    Choosing an Actuator to Compress a Small Spring Precisely

    A 2 lbf (8.90 N) spring over 0.5–0.75 in of travel at 0.025 in steps is a screw problem, not a motor problem: a manual fine-lead screw with a locknut holds the setting with no power. Add a stepper only for remote adjustment, an integrated actuator for automated sequences, or a pneumatic cylinder with an adjustable stop when only endpoint positions matter.

    Motors, gearboxes and drives sizing

  • Selection guide

    Sizing a Vertical Linear Axis for a SCARA Arm

    Size the vertical axis for the complete moving mass, not the 500 g payload. In the worked example an 18.0 kg moving assembly needs 185.6 N of peak thrust and puts a 38.5 N·m moment on the carriage, so a 500 N-class belt-driven axis with a holding brake is the pick once the arm maker approves side mounting on the carriage.

    Linear motion, bearings and guides